Caring for Your Defibrillator
STORING YOUR DEFIBRILLATOR
Always store your defibrillator according to the temperature ranges recommended in
Appendix A
,
"Specifications"
.
CLEANING YOUR DEFIBRILLATOR
WARNING!
Possible fire or explosion.
Do not store this defibrillator in the presence of flammable gases or in direct contact with flammable
material.
CAUTION!
Possible equipment damage.
Do not clean any part of the defibrillator or accessories with bleach, bleach dilution, or phenolic
compounds. Do not use abrasive or flammable cleaning agents. Do not steam, autoclave, or
gas-sterilize the defibrillator or its accessories.
Table 5-1
Cleaning Methods
Item
Cleaning Method
Cleaning Agent
Exterior case, readiness
display, and crevices
Clean with damp sponge or
cloth
Nonabrasive soap and water
Quaternary ammonium compounds
Rubbing (isopropyl) alcohol
Peroxide (peracetic acid) solutions
CHARGE-PAK battery
charger
None
None, dispose of/recycle after use
Electrode Pads
None, do not remove electrode
pads from the packet
None, dispose of/recycle after use
Carrying case
Wipe with damp cloth or sponge
Water