Physio-Control LIFEPAK CR Plus User Manual
Page 33
![background image](/manuals/313531/33/background.png)
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
3-7
©2009-2010 Physio-Control, Inc.
3
U
sing the
Defibrilla
tor
Using the Defibrillator
Defibrillator does not
deliver voice
instructions or beeping
tones after you open it
(turn it on)
Depleted internal battery
•
Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
•
Replace the CHARGE-PAK
battery charger as soon as
possible. After the
OK
symbol
appears on the readiness
display, return the defibrillator
to service.
•
Contact authorized service
personnel.
Speaker system failure
•
Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
•
Contact authorized service
personnel.
The readiness display
is blank
The defibrillator has been turned on
•
Normal condition when the
defibrillator is in use.
Operating temperature is too low or too
high
•
Operate the defibrillator
within 0° to 50°C (32° to
122°F).
LCD not operating properly
•
Contact authorized service
personnel.
Table 3-1
Troubleshooting During Victim Use (Continued)
Problem
Possible Cause
What To Do