LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
A-3
©2009-2010 Physio-Control, Inc.
Ap
pe
nd
ixe
s
Specifications
Charge Ready Time:
Timing
Parameter
200 Joules
360 Joules
Comments
Analysis time
Less than 10
seconds
Less than 10
seconds
Analysis time applies if no
motion is detected. If the
device detects motion,
analysis may be delayed by
up to 10 seconds.
Defibrillator
charge time
Less than 9
seconds
Less than 15
seconds
Defibrillator charge time
applies when the device is
fully charged or has been
discharged up to 15 times.
Power on to
charge ready
time
Less than 26
seconds
Less than 32
seconds
Power on to charge ready
time applies when a
shockable rhythm is detected
and no motion is detected,
and when the device is fully
charged or has been
discharged up to 15 times.
System Recharge
Times:
Recharge times with a fully discharged device:
Able to deliver 6 shocks or provide 42 minutes of operating time after
24 hours of recharge and 20 shocks or 140 minutes of operating time after
72 hours of recharge time with a new CHARGE-PAK battery charger at
temperatures above 15°C (59°F).
Controls:
LID RELEASE/ON-OFF
button—Controls device power.
SHOCK
button (semiautomatic version)—Delivers defibrillation energy.
After electrodes are attached to a victim, an automatic version of the device
delivers a shock, if appropriate, not requiring operator intervention.
Electrical Protection:
Input protected against high voltage defibrillator pulses per
IEC60601-1/EN60601-1. See Figure A-3.
Figure A-3
Defibrillation-protected, Type BF Patient Connection
Safety Classification:
Internally powered equipment. IEC60601-1/EN60601-1.