Physical characteristics accessories – Physio-Control LIFEPAK CR Plus User Manual
Page 57
![background image](/manuals/313531/57/background.png)
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
A-5
©2009-2010 Physio-Control, Inc.
Ap
pe
nd
ixe
s
Specifications
Physical Characteristics
Accessories
CHARGE-PAK Battery Charger
QUIK-PAK Electrode Pads
Height:
10.7 cm (4.2 in)
Width:
20.3 cm (8.0 in)
Depth:
24.1 cm (9.5 in), excluding handle
Weight:
2.0 kg (4.5 lb) with CHARGE-PAK and electrodes
Type:
Li/SO
2
Cl
2
Lithium Sulfuryl Chloride, 11.7V, 1.4 amp-hours.
Replacement:
Replace after each use or when CHARGE-PAK indicator is visible (typically
after 2 years)
Weight:
80.5 grams (0.18 lb)
Pads:
Pacing/defibrillation/ECG electrodes.
Pads Packaging:
User-intuitive, rapid-release QUIK-PAK electrodes allow the electrode pads
to be pre-connected to the device and protected under a top cover.
Pads Shelf Life:
Two years typical.
Electrode Shape:
Oval-rectangular.
Electrode Size:
11.2 cm (4.4 in) × 18.5 cm (7.3 in)
Lead Wire:
1.067 meters (3.5 feet)
Conductive Adhesive
Gel Contact Area:
82 cm
2
(12.8 in
2
)
Maximum Adhesion
Time:
24 hours
Maximum ECG
Monitoring Time:
24 hours
Maximum Number of
Defibrillation Pulses:
50 at 360 joules
Maximum Pacing
Duration:
Up to 12 hours
ECG is received from disposable defibrillation electrodes, standard placement (anterior-lateral), or
anterior/posterior placement.