Multi 3410

Operation

95

ba75863e11

02/2011

4.11.4 Options for data transmission to a PC (USB-B interface)

Via the USB-B interface you can transmit data to a PC. The following
table shows which data are transmitted to the interface in which way:

Note
The following rule applies: With the exception of the menus, shortly
pressing the key generally outputs the display contents to the
interface (displayed measured values, measuring datasets, calibration
records).

4

Set the same transmission data at the connected instrument
(PC):
z Baud rate: to be selected in the range 1200 ... 19200
z Handshake: RTS/CTS
z Set at the PC only:

– Parity: none

– Data bits: 8

– Stop bits: 1

Data

Control

Operation / description

Current
measured values
of all connected
sensors

Manual

z With
z Simultaneously with every

manual storage process
(see section 4.10.1)

Automatic, at
intervals

z With

_

>.

Then you can set the trans-
mission interval

z Simultaneously with every

automatic storage process
(see section 4.10.2)

Stored measured
values

Manual

z Displayed dataset with

after calling up from
the storage

z All datasets with the Output

to RS232/USB function

For details, see section 4.10.3.

calibration
records

Manual

z Displayed calibration record

with

z All calibration records with

_

>

For details, see section 4.11.

Automatic

z At the end of a calibration

procedure