Multi 3410

Operation

41

ba75863e11

02/2011

4.6.3

Settings for pH measurements

Overview

The following settings are possible for pH measurements:

z Calibration record (display)

z Calibration data storage (display)

z Calibration interval

z Buffers for calibration

z Single-point calibration

z Unit for slope

z Resolution

Settings

The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the

key. After completing the settings, switch to the mea-
sured value display with .
Default settings are printed in bold.

Menu item

Possible setting

Description

Calibration /Calibration record

-

Displays the calibration record of the last cal-
ibration

Calibration /Calibration data
storage

-

Displays the last calibration records (max.
10)

Calibration / Calibration data
storage
/ Copy to USB flash
drive

-

Outputs the calibration data storage to a con-
nected USB flash drive

Calibration / Calibration data
storage
/ Output to RS232/USB

-

Outputs the calibration data storage to the
interface

Calibration /Buffer

TEC
NIST/DIN
ConCal
...

Buffer sets to be used for pH calibration.
For further buffers and details, see section
4.6.4.

Calibration /Single-point cali-
bration

yes
no

Quick calibration with 1 buffer

Calibration /Calibration interval

1 ... 7 ... 999 d

Calibration interval for the IDS pH sensor (in
days).

The meter reminds you to calibrate regularly
by the flashing sensor symbol in the measur-
ing window.