3 settings for ph measurements, Settings for ph measurements – Xylem Multi 3410 User Manual
Page 41

Multi 3410
Operation
41
ba75863e11
02/2011
4.6.3
Settings for pH measurements
Overview
The following settings are possible for pH measurements:
z Calibration record (display)
z Calibration data storage (display)
z Calibration interval
z Buffers for calibration
z Single-point calibration
z Unit for slope
z Resolution
Settings
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the
key. After completing the settings, switch to the mea-
sured value display with
Default settings are printed in bold.
Menu item
Possible setting
Description
Calibration /Calibration record
-
Displays the calibration record of the last cal-
ibration
Calibration /Calibration data
storage
-
Displays the last calibration records (max.
10)
Calibration / Calibration data
storage / Copy to USB flash
drive
-
Outputs the calibration data storage to a con-
nected USB flash drive
Calibration / Calibration data
storage / Output to RS232/USB
-
Outputs the calibration data storage to the
interface
Calibration /Buffer
TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH calibration.
For further buffers and details, see section
4.6.4.
Calibration /Single-point cali-
bration
yes
no
Quick calibration with 1 buffer
Calibration /Calibration interval
1 ... 7 ... 999 d
Calibration interval for the IDS pH sensor (in
days).
The meter reminds you to calibrate regularly
by the flashing sensor symbol in the measur-
ing window.