02/2011

4.9.4

Settings for IDS conductivity sensors

Overview

The following settings are possible for IDS conductivity sensors:

z Calibration record (display)

z Calibration data storage (display)

z Calibration interval

z Reference temperature

z Temperature compensation

z TDS factor

z Unit of the temperature

Settings

The settings are made in the menu for the measured parameter, con-
ductivity. To open the settings, display the required parameter in the
measured value display and press the

key. After
completing the settings, switch to the measured value display with
.
The possible settings are individually displayed for each sensor. Below
the setting menu is displayed for two IDS sensors (TetraCon 925,
LR325/01).
Default settings are printed in bold.

Setting menu of

TetraCon 925

Menu item

Possible
setting

Description

Calibration /Cali-
bration record

-

Displays the calibration record of
the last calibration

Calibration /Cali-
bration data stor-
age

-

Displays the last calibration
records (max. 10)

Calibration /
Calibration data
storage
/ Copy to
USB flash drive

-

Outputs the calibration data sto-
rage to a connected USB flash
drive

Calibration /Cali-
bration interval

1 ... 150 ...
999 d

Calibration interval for the IDS
conductivity sensor (in days).

The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
window.