beautypg.com

Xylem Multi 3410 User Manual

Page 51

background image

Multi 3410

Operation

51

ba75863e11

02/2011

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.

7

Wait for the end of the measurement with stability control or
accept the calibration value with

.
The calibration display for the next buffer appears (voltage dis-
play).

8

If necessary, finish the calibration procedure as a single-point
calibration with .
The calibration record is displayed.

22.09.2009 08:00