Xylem Multi 3410 User Manual
Page 68

Operation
Multi 3410
68
ba75863e11
02/2011
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.
You can start a measurement with Stability control manually at any
time, irrespective of the setting for automatic Stability control (see sec-
tion 4.4.3) in the System menu.
Note
You can prematurely terminate the Stability control function manually
with at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are not output to
the interface.
2
Release the frozen measured value again with
The HOLD function is switched off.
The [HOLD] status display disappears.
1
Freeze the measured value with
The [HOLD] status indicator is displayed.
2
Using , activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
3
Using , start a further measurement with Sta-
bility control.
or
Release the frozen measured value again with
The display switches to the measured value display.
The [AR][HOLD] status display disappears.