5 calibration interval, Calibration interval – Xylem Multi 3410 User Manual

Page 49

background image

Multi 3410

Operation

49

ba75863e11

02/2011

Preparatory activities

Perform the following preparatory activities when you want to calibrate:

4.6.5

Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is
replaced by the QSC color scale (see section 4.6.10).

After the specified calibration interval has expired, the sensor symbol
or the QSC color scale flashes. It is still possible to measure.

Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.

Setting the calibration

interval

The calibration interval is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):

1

Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.

2

Keep the buffer solutions ready.

1

Open the menu for measurement settings with

ENTER>.

2

In the Calibration / Calibration interval menu, set the calibration
interval with <

S>>.

3

Confirm the setting with

.

4

Quit the menu with .