beautypg.com

5 d.o. calibration, D.o. calibration – Xylem Multi 3420 User Manual

Page 74

background image

Operation

Multi 3420

74

ba75864e11

02/2011

4.9.5

D.O. calibration

Why calibrate?

D.O. sensors age. This changes the slope of the D.O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.

Note
The FDO

®

925 D.O. sensor ages so little it does not have to be regu-

larly calibrated.
To detect changes of the sensor as early as possible, the FDO

®

Check

procedure can be useful (see section 4.9.4).

When to calibrate?

z If your evaluation of the FDO

®

Check procedure suggests a calibra-

tion

z When the calibration interval has expired

z If your accuracy requirements are especially high

z Routinely within the scope of company quality assurance measures.

Calibration procedures

Calibration in water vapor-saturated air.
To calibrate the FDO

®

925 sensor, use the check and storage vessel,

FDO

®

Check.

Stability control

(AutoRead)

In calibration, the Stability control function (AutoRead) is automatically
activated.

Display calibration data

and output to interface

You can have the data of the last calibration displayed (see section
4.10.6)
. Subsequently, you can transmit the displayed calibration data
to the interface, e. g. to a PC, with the key.

Note
The calibration record is automatically transmitted to the interface after
calibrating.

Sample record

Calibration evaluation

After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.

Multi 3420
Ser. no. 10139695

CALIBRATION Ox
Calibration date

30.03.2010 16:13:33

FDO 925
Ser. no. 10146858

SC-FDO 925

10158765

Relative slope

0.98

Sensor

+++