9 dissolved oxygen, 1 general information, Dissolved oxygen 4.9.1 – Xylem Multi 3420 User Manual

Page 66: General information

background image

Operation

Multi 3420

66

ba75864e11

02/2011

4.9

Dissolved oxygen

4.9.1

General information

You can measure the following parameters:

z D.O. concentration

z D.O. saturation index ("D.O. saturation")

z D.O. partial pressure

Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This facilitates interference-free measurements also in
the following cases:

z Measurement in grounded test samples

z Measurement in one test sample with several sensors connected to

one Multi 3420

Temperature

measurement

IDS D.O. sensors measure the temperature with a temperature sensor
integrated in the IDS sensor.

Preparatory activities

Perform the following preparatory activities when you want to measure:

Note
The FDO

®

925 needs to be calibrated in special cases only. Regular

FDO

®

Checks are sufficient.

1

Connect the D.O. sensor to the meter.
The D.O. measuring window is displayed.

2

Check or calibrate the meter with the sensor.