beautypg.com

Xylem Multi 3420 User Manual

Page 48

background image

Operation

Multi 3420

48

ba75864e11

02/2011

Sample record

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.

Note
For pH IDS sensors, you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 4.7.10).

Multi 3420
Ser. no. 09250023

CALIBRATION

pH

Calibration date

31.07.2009 16:13:33

SenTix 940
Ser. no. B092500013

TEC
Buffer 1

4.01

Buffer 2

7.00

Buffer 3

10.01

Voltage

1

184.0 mV 24.0 °C

Voltage

2

3.0 mV

24.0 °C

Voltage 3

-177.0 mV 24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV

Sensor

+++

etc...

Display

Calibration
record

Zero point
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the IDS sensor according to the
sensor operating manual

Error

Error

< -30 or
> 30

... -62 or
... -50

Eliminate the error according to chapter
6 W

HAT

TO

DO

IF

...