Multi 3420

Operation

41

ba75864e11

02/2011

Note
You can prematurely terminate the Stability control function manually
with

at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.

Criteria for a stable

measured value

The Stability control function checks whether the measured values are
stable within the monitored time interval.

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

3

Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back
to the previous indication.

Measured
parameter

Time interval

Stability in the time interval

pH value

15 seconds

Δ

pH: Better than 0.01

Temperature

15 seconds

Δ

T (° C): Better than 0.02