Xylem Multi 3420 User Manual
Page 47

Multi 3420
Operation
47
ba75864e11
02/2011
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
Note
You can display the slope in the units, mV/pH or % (see section 4.7.3).
Stability control
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).
Calibration record
The new calibration values are displayed when a calibration procedure
is completed.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see section
4.7.8). Subsequently, you can transmit the displayed calibration data to
the interface, e. g. to a PC, with the
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Determined
values
Displayed calibration data
1-point
Asy
z Zero point = Asy
z Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
z Zero point = Asy
z Slope = Slp.
3-point to
5-point
Asy
Slp.
z Zero point = Asy
z Slope = Slp.
The calibration line is calculated
by linear regression.