beautypg.com

3 settings for orp measurements, Settings for orp measurements – Xylem Multi 3420 User Manual

Page 65

background image

Multi 3420

Operation

65

ba75864e11

02/2011

Criteria for a stable

measured value

The Stability control function checks whether the measured values are
stable within the monitored time interval.

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

4.8.3

Settings for ORP measurements

Overview

The following settings are possible for ORP measurements:

z Resolution

z Automatic stability control

Settings

The settings are made in the menu for measuring settings of the ORP
measurement. To open the settings, display the required parameter in
the measured value display and press the

key. After
completing the settings, switch to the measured value display with
.
Default settings are printed in bold.

3

Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back
to the previous indication.

Measured param-
eter

Time interval

Stability in the time inter-
val

ORP voltage

15 seconds

Δ

mV: Better than 0.3

Menu item

Possible set-
ting

Description

mV resolution

0.1
1

Resolution of the mV dis-
play

Reset

-

Resets all sensor settings to
the delivery condition (see
section 4.13.1)