4.7.4

pH calibration

Why calibrate?

During the operation of an IDS pH sensor, the zero point (asymmetry)
and
slope of the sensor change with time.As a result, an inexact measured
value is displayed. Calibration determines the current values of the
zero point and slope of the pH sensor and stores them in the measuring
instrument.
Thus, you should calibrate at regular intervals.

When do you have to

calibrate?

z When the calibration interval has expired

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into consideration
during the calibration.

No.

Buffer set *

pH values

at

1

ConCal

Any

Any

2

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1.679
4.006
6.865
9.180

12.454

25 °C

3

TEC
WTW Technical buffers

2.000
4.010
7.000

10.011

25 °C

4

Merck 1*

4.000
7.000
9.000

20°C

5

Merck 2 *

1.000
6.000
8.000

13.000

20°C

6

Merck 3 *

4.660
6.880
9.220

20°C

7

Merck 4 *

2.000
4.000
7.000

10.000

20°C