Multi 3430

Operation

89

ba75865e11

02/2011

4.11.1 Manual storage

You can transmit a measurement dataset to the data storage as fol-
lows. The dataset is at the same time output to the interface:

If the storage is full

When all storage locations are occupied, it is not possible to continue
storing. Then you can e.g. transmit the stored data to a PC or a USB
flash drive (see section 4.11.3) and subsequently erase the storage
(see section 4.11.4).

4.11.2 Automatic storage at intervals

The storage interval (Interval) determines the time interval between
automatic storage processes. Each storage process transmits the cur-
rent dataset to the interface at the same time.

Configuring the

automatic storage

function

1

Press the key shortly.
The menu for manual storage appears.

2

If necessary, change and confirm the ID number (1 ... 10000)
with <

S>> and

.

The dataset is stored. The meter switches to the measured
value display.

30.09.2009 11:24:16
pH 7.000 24.8 °C AR +++

ID number:

1

Continue

Manual data storage 4 From 500

22.09.2009 08:00

1

Press the

_

> key.

The menu for automatic storage appears.