10 conductivity, 1 general information, 2 measuring – Xylem Multi 3430 User Manual
Page 78: 1 general information 4.10.2 measuring

Operation
Multi 3430
78
ba75865e11
02/2011
4.10
Conductivity
4.10.1 General information
You can measure the following parameters:
z Conductivity
z Specific resistance
z Salinity
z Total dissolved solids (TDS)
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This facilitates interference-free measurements also in
the following cases:
z Measurement in grounded test samples
z Measurement in one test sample with several sensors connected to
one Multi 3430
Temperature
measurement
IDS sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
Preparatory activities
Perform the following preparatory activities when you want to measure:
4.10.2 Measuring
You can carry out conductivity measurements as follows:
1
Connect an IDS conductivity sensor to the meter.
The conductivity measuring window is displayed.
The Measuring cell and cell constant for the connected IDS
conductivity sensor are automatically taken over.
1
Perform the preparatory activities according to section 4.10.1.
2
Immerse the IDS conductivity sensor in the test sample.