beautypg.com

3 dissolved oxygen, Dissolved oxygen – Xylem Multi 3430 User Manual

Page 106

background image

What to do if...

Multi 3430

106

ba75865e11

02/2011

No stable measured

value

Obviously incorrect

measured values

6.3

Dissolved oxygen

Error message,

OFL

Cause

Remedy

IDS pH sensor:

– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if neces-

sary

IDS pH sensor + test sample:

– Conductivity too low

– Use suitable IDS pH sensor

– Temperature too high

– Use suitable IDS pH sensor

– Organic liquids

– Use suitable IDS pH sensor

Cause

Remedy

IDS pH sensor:

– IDS pH sensor unsuitable

– Use suitable IDS sensor

– Temperature difference

between buffer and test sample
too great

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Measured value outside the

measuring range

– Select a suitable measuring

medium