Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.

7

Wait for the end of the measurement with stability control or
accept the calibration value with

.
The calibration display for the next buffer appears (voltage dis-
play).

8

If necessary, finish the calibration procedure as a single-point
calibration with .
The calibration record is displayed.