Multi 3430

Operation

71

ba75865e11

02/2011

FDO Check /
Check interval

1 ... 60 ...
999 d

Interval for the FDO Check (in days).

The FDO Check status display indica-
tor in the measuring window reminds
you to check the sensor regularly.

Sal automatic

on
off

Automatic salt content correction for
concentration measurements. The
salinity value is taken from an IDS
conductivity sensor.

This menu item is only available if an
IDS conductivity sensor is connected.

Salinity from
channel

▭▭▬

▭▬▭

▬▭▭

Selects the channel (sensor), provi-
ding the salinity value.

This menu item is only available if two
IDS conductivity sensors are con-
nected.

Sal correction

on
off

Manual salt content correction for con-
centration measurements

Salinity

0.0 ...
70.0

Salinity or salinity equivalent for the
salt content correction.

This function is only available for con-
centration measurements if the man-
ual salt content correction is switched
on.

Response time
t90

30 ... 300

Response time of the signal filter (in
seconds).

A signal filter in the sensor reduces the
fluctuation range of the measured
value. The signal filter is characterized
by the response time t90. This is the
time after which 90 % of a signal
change is displayed.

This menu item is only available if sen-
sor and meter support this function.
Firmware updates for the IDS sensor
and the meter are provided on the
internet (see chapter 9).

Reset

-

Resets all sensor settings to the deliv-
ery condition (see section 4.13.1)

Menu item

Possible
setting

Description