4 conductivity, Conductivity – Xylem Multi 3430 User Manual

Page 107

background image

Multi 3430

What to do if...

107

ba75865e11

02/2011

Error message,

Error

6.4

Conductivity

Error message,

OFL

Error message,

Error

Cause

Remedy

– Sensor contaminated

– Clean the sensor

– Measured temperature outside

the operating conditions
(OFL/UFL displayed instead of
temperature value)

– Observe the temperature

range for the measuring
medium

– Defective sensor

– Replace sensor

Note
More information, as well as notes on cleaning and replacement of
sensors, are given in the documentation of your sensor.

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable IDS conductiv-

ity sensor

Cause

Remedy

– Sensor contaminated

– Clean sensor and replace it

if necessary

– Calibration solution not suitable

– Check calibration solutions

Note
More information, as well as notes on cleaning and replacement of
sensors, are given in the documentation of your sensor.