4.7.8

Displaying calibration records

The calibration data can be displayed and then output to the interface.

Displaying the

calibration record

The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the

_

> key.

The calibration records of the last 10 calibrations are to be found in the
menu, Calibration / Calibration data storage / Display. To open the Cal-
ibration
menu, press the

key in the measured value
display.

Menu item

Setting/
function

Description

Calibration /
Calibration data stor-
age
/ Display

-

Displays the calibration
records.

Further options:
z Scroll through the calibra-

tion records with
<

W>>.

z Output the displayed cali-

bration record to the inter-
face with .

z Output all calibration

records to the interface
with

_

>.

z Quit the display with

or

ENTER>.

z Switch directly to the mea-

sured value display with
.

Calibration /
Calibration data stor-
age
/ Copy to USB
flash drive

-

Outputs the calibration data
storage to a connected USB
flash drive

Calibration /
Calibration data stor-
age
/ Output to
RS232/USB

-

Outputs the calibration data
storage to the interface