3 calibration settings and calibration data, Calibration settings and calibration data – Xylem ProLab 3000 LABORATORY pH METER User Manual
Page 85

ProLab 3000
pH
85
ba75645e02
01/2010
7.3.3
Calibration settings and calibration data
In the Calibration menu, you can
z set the calibration interval,
z define the calibration type and
z view the calibration data.
Menu item
Setting
Description
Calibration
– pH calibration type
– AutoCal
( )
Calibration with commercial buffer sets.
The buffer set is selected with the next
menu item (Buffer sets)
– Buffer sets
DIN ( )
TEC
...
Buffer sets to be used for pH calibration.
For details, see page 72.
– VariCal
( )
Calibration with any buffer solutions
– Slope in
mV/pH ( )
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.16 mV/pH ([determined slope/Nernst
slope ] x 100).
– Zero point in
mV ( )
pH
Unit for the zero point.
– Calibration interval...
1 ... 999
Calibration interval... for the pH electrode
(in days).
The meter reminds you to calibrate regu-
larly by the red frame around the CalClock
in the measuring window.
– Calibration history...
Displays the calibration history of the last
calibrations.