Xylem ProLab 3000 LABORATORY pH METER User Manual
Page 106

Ion concentration
ProLab 3000
106
ba75645e02
01/2010
Continuing with two-
point calibration
9
Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The ISE standard input window for the entry of the concentra-
tion of the second standard solution pops up.
10
Select a value from the ISE-Std list.
or
Enter a value and, if necessary, an exponent with <0...9>.
or
Cancel the calibration with [Cancel].
Note
To enter a concentration in exponential representation such as
1.00 E2 mg/l:
Enter the mantissa with <0...9> (here: 1.00).
Enter the exponential sign (E) with <- .><- .>.
Enter the exponent with <0...9> (here: 2).
11
When measuring without temperature sensor:
z Measure the temperature of the standard solution using a
thermometer.
z Open the setting window for the temperature value with
[Man. temp.].
z Enter the temperature and confirm with [OK].
12
Thoroughly rinse the electrode with distilled water.
13
Immerse the electrode in standard solution 2.
14
Start the measurement with [OK].
The measured value is checked for stability (Stability control).
15
Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The ISEX calibration window for the next standard solution
pops up and the Finish with 2-point calibration infotext is dis-
played.