Xylem ProLab 3000 LABORATORY pH METER User Manual
Page 166

What to do if...
ProLab 3000
166
ba75645e02
01/2010
No stable measured
value
Obviously incorrect
measured values
– Electrode broken
– Replace the electrode
Buffer solutions:
– Incorrect buffer solutions
– Select the suitable buffer set
for the buffers used
or
– Replace the buffer solutions
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Replace the buffer solutions
Cause
Remedy
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if neces-
sary
Electrode + test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Use a suitable electrode
– Organic liquids
– Use a suitable electrode
Cause
Remedy
pH electrode:
– Not connected
– Connect the electrode
– Cable broken
– Replace the electrode
– pH electrode unsuitable
– Use a suitable electrode
– Temperature difference
between buffer and test sam-
ple too high
– Adjust temperature of buffer
or sample solutions