Program installation
Program start
Firmware update
After switching the meter off and on you can check whether the meter
has taken over the new software version.
11
Connect the ProLab 3000 to a USB interface (virtual COM port)
of the PC with the aid of the USB interface cable.
or
Connect the ProLab 3000 to a serial interface (COM port) of
the PC with the aid of the Z390 interface cable.
1
With the installation program,
"Install_Update_ProLab3000_4000_Vx_yy_English.exe",
install the firmware update program on a PC.
2
Start the "Update_ProLab3000_4000" program from the Win-
dows start menu.
3
If necessary, change the language in the language menu.
4
Start the updating procedure in the firmware update program
with [OK].
5
Follow the instructions of the firmware update program.
During the programming process, a corresponding message
and a progress bar is displayed.
The programming process takes up to 15 minutes.
A terminatory message is displayed after a successful pro-
gramming process. The firmware update is completed.
6
Disconnect the ProLab 3000 from the PC.
The ProLab 3000 is ready for operation.