background image

pH 3110

Technical data

43

ba75791e02

11/2008

7

Technical data

7.1

General data

Dimensions

approx. 240 x 190 x 80 mm

Weight

approx. 0.4 kg

Mechanical structure

Type of protection

IP 67

Electrical safety

Protective class

III

Test certificates

CE

Ambient

conditions

Storage

- 25 °C ... + 65 °C

Operation

-10 °C ... + 55 °C

Climatic class

2

Power

supply

Batteries

4 x 1.5 V alkali-manganese batteries,
type AA

Rechargeable
batteries

4 x 1,2 V NiMH rechargeable batteries,
type AA
(no charging function)

Operational life

Approx. 2500 h operating hours (batteries)

Sensor input

Input resistance

> 5 * 10

12

ohm

Input current

< 1 * 10

-12

A

Guidelines

and norms used

EMC

EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A

Meter safety

EC directive 2006/95/EC
EN 61010-1

Climatic class

VDI/VDE 3540

IP protection

EN 60529