pH 3110

What to do if...

41

ba75791e02

11/2008

Obviously incorrect

measured values

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

pH combination electrode:

– pH combination electrode

unsuitable

– Use suitable combination

electrode

– Temperature difference

between buffer and test sample
too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the and

key

simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

During the display test,
display the software version
with .