pH 3110

Operation

25

ba75791e02

11/2008

Note
Here you can cancel the calibration procedure with . This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. The sensor symbol shows the evaluation of the
single-point calibration. Pressing displays the value of the
slope (SLO).

2

If necessary, set the temperature of the first buffer solution with
<

▲><▼>.

3

Immerse the pH combination electrode in the first buffer
solution (pH 7.0 ± 0.5 for two-point calibration).

4

Press the key.
The measured pH value appears on the display.

5

When the measured value is stable,
using <

▲><▼>, set the measured value to the nominal pH

value of the buffer solution (at the current temperature).

6

Press the key. SLO appears on the display.

8

4

2

Y

S

A

°C

TP

ConCal

8

4

2

O

L

s

°C

TP

ConCal