pH 3110

What to do if...

39

ba75791e02

11/2008

6

What to do if...

Error message

OFL, UFL

Error message

CalError

Cause

Remedy

pH combination electrode:

– Measured value outside the

measuring range

– Use suitable combination

electrode

– Air bubble in front of the

diaphragm

– Remove air bubble

– Air in the diaphragm

– Extract air or moisten

diaphragm

– Cable broken

– Replace combination

electrode

– Gel electrolyte dried out

– Replace combination

electrode

Cause

Remedy

pH combination electrode:

– The values determined for zero

point and slope of the
combination electrode are
outside the allowed limits.

– Recalibrate

– Diaphragm contaminated

– Clean diaphragm

– Combination electrode broken

– Replace combination

electrode

Buffer solutions

– Incorrect buffer solutions

– Change calibration

procedure

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions