2 measurement settings, Measurement settings, Fu b – Xylem PH3110 FIELD PH METER User Manual

Page 30

background image

Operation

pH 3110

30

ba75791e02

11/2008

4.5.2

Measurement settings

These settings apply to calibration and measurement (the default
setting is printed in bold).

Display during

calibration

(bUFF)

Display during calibration (BUFFER)

pH (buffer nominal value),
U (combination electrode
voltage)

Unit of the value for the slope (SLO)

mV/pH, %

Unit of the value for the zero point
(ASY)

mV, pH

Measured value resolution, pH

0.000 , 0.00 , 0.0

Measured value resolution, U

0.0 , 0

Temperature unit (UnIt)

°C, °F

Calibration interval (Int.C)

0 ... 7 ... 999 d

1

With

_

> open the menu for measurement and calibration

settings.
The first setting is displayed.

2

Using <

▲><▼>, select the pH or U display during calibration.

3

Confirm with .
SLO, the unit of the value for the slope (mV/pH or %) is
displayed.

H

p

F

F

U

b

pH

AutoCal DIN

AutoCal TEC