2 automatic calibration (autocal), Automatic calibration (autocal), 1t c – Xylem PH3110 FIELD PH METER User Manual

Page 22: Id c, 2t c, 2d c

background image

Operation

pH 3110

22

ba75791e02

11/2008

4.3.2

Automatic calibration (AutoCal)

For the AutoCal TEC and AutoCal DIN procedures, use one to three
buffer solutions of the relevant buffer set (Technical buffers or DIN
buffers) in any order.

Note
The steps 2 and 6 are not necessary if you use a temperature sensor.

1

Press repeatedly until the AutoCal TEC or AutoCal DIN
function display indicator appears.

8

4

2

1

T

C

°C

pH

AutoCal TEC

TP

8

4

2

I

D

C

°C

pH

AutoCal DIN

TP

2

If necessary, set the temperature of the buffer solution with
<

▲><▼>.

3

Immerse the pH combination electrode in the first buffer
solution.

4

Start the measurement with .
The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. As soon as a stable
value is recognized, Ct2 or Cd2 appears.

8

4

2

2

T

C

°C

pH

AutoCal TEC

TP

8

4

2

2

D

C

°C

pH

AutoCal DIN

TP