pH 3110

Operation

23

ba75791e02

11/2008

Note
Here you can cancel the calibration procedure with . This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. Pressing displays the value of the slope
(SLO).

Continuing with two-

point calibration

Note
Here you can cancel the calibration procedure with . This
corresponds to a two-point calibration. The value of the zero point
(ASY) is displayed. Pressing displays the value of the slope
(SLO).

Continuing with three-

point calibration

5

Thoroughly rinse the combination electrode with distilled water.

6

If necessary, set the temperature of the second buffer solution
with <

▲><▼>.

7

Immerse the combination electrode in the second buffer
solution.

8

Press the

key.

The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. As soon as a stable
value is recognized, Ct3 or Cd3 appears.

8

4

2

3

T

C

°C

pH

AutoCal TEC

TP

8

4

2

3

D

C

°C

pH

AutoCal DIN

TP

9

Thoroughly rinse the combination electrode with distilled water.

10

If necessary, set the temperature of the third buffer solution
with <

▲><▼>.

11

Immerse the combination electrode in the third buffer solution.