Cond 3210

What to do if...

45

ba75794e01

01/2009

6

What to do if...

Error message,

OFL

Error message,

Error

Sensor symbol flashes

Display

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Calibration solution not suitable

– Check calibration solutions

Cause

Remedy

– Cleaning interval expired

– Recalibrate the measuring

system

Cause

Remedy

– batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the and
key
simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

Open the menu,

_

>[Menu] / Storage &

config / System / Service
information
. The instrument
data is displayed.