Xylem COND3210 HANDHELD CONDUCTIVITY METERS User Manual
Page 29

Cond 3210
Operation
29
ba75794e01
01/2009
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 26) in the
Measurement menu.
Note
You can prematurely terminate the Stability control function manually
with
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
1
Freeze the measured value with
The [HOLD] status indicator is displayed.
2
Using
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
3
Using
control.
or
Release the frozen measured value again with
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ӕ
10 seconds
Δ
ӕ
: better than
0.5% of measured value
Temperature
10 seconds
Δ
T (° C): Better than 0.02