6 reset, 1 resetting the measurement settings, Reset 4.6.1 – Xylem COND3210 HANDHELD CONDUCTIVITY METERS User Manual
Page 40: Resetting the measurement settings

Operation
Cond 3210
40
ba75794e01
01/2009
4.6
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1
Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the
default settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the
shortly.
Setting
Default settings
Cal. interval
150 d
Measured parameter
ӕ
Cell constant (c)
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of
the linear temperature
compensation
2.000 %/K
TDS factor
1.00
Stability control
On
Temperature unit
°C