Operation

Cond 3210

40

ba75794e01

01/2009

4.6

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.6.1

Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

The following settings for conductivity measurements are reset to the
default settings with the Reset function:

The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the [Menu] key
shortly.

Setting

Default settings

Cal. interval

150 d

Measured parameter

ӕ

Cell constant (c)

0.475 cm

-1

(calibrated)

0.475 cm

-1

(set up)

Temperature compensation

nLF

Reference temperature

25 °C

Temperature coefficient (TC) of
the linear temperature
compensation

2.000 %/K

TDS factor

1.00

Stability control

On

Temperature unit

°C