4.4.3

Temperature compensation

The calculation of the temperature compensation is based on the
preset reference temperature, 20 °C or 25 °C. It appears on the display
as Tr20 or Tr25.

You can select one of the following temperature compensation
methods:

z Nonlinear temperature compensation (nLF)

according to EN 27 888

z Linear temperature compensation (lin)

with adjustable coefficients of 0.000 ... 3.000 %

/

K

z No temperature compensation (off)

Note
The reference temperature and temperature compensation are set in
the conductivity measuring menu (see section 4.4.4).

Application tips

Select the following temperature compensations given in the table
according to the respective test sample:

Test sample

Temperature compensation

Display
indicator

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous
solutions

lin
Set linear temperature
coefficient
0.001 ... 3.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT

Sal, nLF