background image

Operation

Cond 3210

32

ba75794e01

01/2009

Measurement /
Measuring cell /
Type

Measuring cell used

Cal

Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution.
Calibration ranges:
0.450 to 0.500 cm

-1

and

0.800 to 0.880 cm

-1

The currently valid cell
constant is displayed in the
status line.

LR325/01

LR 325/01 measuring cell,
nominal cell constant
0.100 cm

-1

.

The cell constant can be
adjusted in the range from
0.090 to 0.110 cm

-1

.

LR325/001

LR 325/001 measuring cell,
nominal cell constant
0.010 cm

-1

.

The cell constant is
permanently set.

man

Any measuring cells with
freely adjustable cell
constants in the range from
0.250 to 25.000 cm

-1

.

Measurement /
Measuring cell /
Cell const. man

0.250 to
25.000 cm

-1

Display and setting option of
the cell constant of any
measuring cells (man).

Measurement /
Measuring cell /
Cell const. LR325/01

0.090 to
0.110 cm

-1

Display and setting option of
the cell constant of the
LR 325/01 measuring cell.

Measurement /
Temp. comp. (TC) /
Method

nLF
lin
Off

Procedure for temperature
compensation (see section
4.4.3)
.

This setting is only available
for the measured
parameters,

ӕ

und

ρ

.

Menu item

Possible
setting

Description