Xylem COND3210 HANDHELD CONDUCTIVITY METERS User Manual
Page 32
Operation
Cond 3210
32
ba75794e01
01/2009
Measurement /
Measuring cell /
Type
Measuring cell used
Cal
Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution.
Calibration ranges:
0.450 to 0.500 cm
-1
and
0.800 to 0.880 cm
-1
The currently valid cell
constant is displayed in the
status line.
LR325/01
LR 325/01 measuring cell,
nominal cell constant
0.100 cm
-1
.
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm
-1
.
LR325/001
LR 325/001 measuring cell,
nominal cell constant
0.010 cm
-1
.
The cell constant is
permanently set.
man
Any measuring cells with
freely adjustable cell
constants in the range from
0.250 to 25.000 cm
-1
.
Measurement /
Measuring cell /
Cell const. man
0.250 to
25.000 cm
-1
Display and setting option of
the cell constant of any
measuring cells (man).
Measurement /
Measuring cell /
Cell const. LR325/01
0.090 to
0.110 cm
-1
Display and setting option of
the cell constant of the
LR 325/01 measuring cell.
Measurement /
Temp. comp. (TC) /
Method
nLF
lin
Off
Procedure for temperature
compensation (see section
4.4.3).
This setting is only available
for the measured
parameters,
ӕ
und
ρ
.
Menu item
Possible
setting
Description