Cond 3210

Operation

31

ba75794e01

01/2009

4.4.4

Settings for conductivity measuring cells

Overview

The following settings are possible for conductivity measuring cells:

z Calibration record (display)

z Calibration interval

z Measuring cell/cell constant

z Reference temperature

z Temperature compensation

z TDS factor

z Unit of the temperature

z Automatic Stability control

Settings

The settings are made in the measuring menu of the conductivity
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the /[menu] or
key shortly. After completing the settings, switch to the
measured value display with .

Menu item

Possible
setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.