Cond 3210
Operation
31
ba75794e01
01/2009
4.4.4
Settings for conductivity measuring cells
Overview
The following settings are possible for conductivity measuring cells:
z Calibration record (display)
z Calibration interval
z Measuring cell/cell constant
z Reference temperature
z Temperature compensation
z TDS factor
z Unit of the temperature
z Automatic Stability control
Settings
The settings are made in the measuring menu of the conductivity
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the
measured value display with
Menu item
Possible
setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.