6 displaying calibration records, Displaying calibration records, Section 4.4.6) – Xylem COND3210 HANDHELD CONDUCTIVITY METERS User Manual

Page 35

background image

Cond 3210

Operation

35

ba75794e01

01/2009

Determining the cell

constant

For this calibration procedure, the Measuring cell setting must be set to
cal in the measuring menu. Proceed as follows to determine the cell
constant:

4.4.6

Displaying calibration records

Displaying the

calibration record

The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the

_

> key.

1

Connect a conductivity measuring cell to the measuring
instrument.

2

In the measured value display, select the conductivity
parameter with .

3

Start the calibration with .
The cell constant that was calibrated last is displayed.

4

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

5

Start the measurement with .
The [AR] status indicator is displayed. The measured
parameter flashes.

6

Wait for the end of the AutoRead measurement or accept the
calibration value with .
The calibration record is displayed.

7

Using /[Continue] or , switch to the measured
value display.

ӕ

0.839

1/cm

25.1

°C

[Tr25][nLF][ 0.835 1/cm]

22.09.2008

08:00