Pressure resistance
Max. allowed overpressure (sensor including connection cable):
The SensoLyt
®
700 IQ (SW) meets the requirements according to
article 3(3) of the directive, 97/23/EC ("pressure equipment directive").
Immersion depth
CAUTION
The pressure resistance of the operable pH/ORP armature can be
reduced by the pressure resistance of the combination electrode.
When selecting the combination electrode make sure it is suitable
for the intended pressure and temperature range.
Type of protection
Operating position
any
Fields of application
with installed combination electrode,
SensoLyt
®
SEA, DWA, PtA
10
6
Pa (10 bar) *
with installed combination electrode,
SensoLyt
®
ECA
6 x10
5
Pa (6 bar) *
with installed combination electrode,
SensoLyt
®
SEA-HP
10
6
Pa (10 bar) **
* temperature dependent (see safety instruction below)
** in the entire temperature range
with installed combination electrode,
SensoLyt
®
SEA, DWA, PtA
min. 40 mm; max. 100 m *
with installed combination electrode,
SensoLyt
®
ECA
min. 40 mm; max. 60 m *
with installed combination electrode,
SensoLyt
®
SEA-HP
min. 40 mm; max. 100 m **
* temperature dependent (see safety instruction below)
** in the entire temperature range
Sensor with installed combination
electrode and including SACIQ (SW)
sensor connection cable
IP 68, 10 bar (10
6
Pa)
Sensor plug head connector without
sensor connection cable (sensor with
installed combination electrode)
IP 67
SensoLyt 700 IQ
Stationary measurements in water/
wastewater applications
SensoLyt 700 IQ SW
Stationary measurements in seawater
and brackish water, aquaculture