SensoLyt
®
700 IQ (SW)
Measuring / Operation
4 - 5
ba76006e01
01/2012
4.2.4
Calibration with CAL CON 1P
Messages displayed during the single-point calibration with
CAL CON 1P
Display
Explanation
* Have any buffer solution ready.
You can use any buffer solution
the pH value of which is known at
the current temperature.
* Rinse the sensor.
* Immerse the sensor in the buffer
solution.
* Wait for a stable measured
value.
Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.
* Enter the pH value of the buffer
solution.
Enter the pH value with
d
.
Confirm entered data and
continue calibration
Confirm with
g
. The
measurement of the buffer
solution begins.
Successfully calibrated.
End of the CAL CON 1P
calibration.
The values determined for Slope
and Asymmetry potential are
displayed.
The calibration is complete.
Confirm with
g
.
The display returns to the
measured value display.
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)