4 sensolyt® 700 iq (sw) setting table, Sensolyt – YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual
Page 16
![background image](/manuals/310719/16/background.png)
Commissioning
SensoLyt
®
700 IQ (SW)
3 - 6
ba76006e01
01/2012
3.4
SensoLyt
®
700 IQ (SW) setting table
Setting
Selection/values
Explanation
Measuring mode
mV
pH
Unit of the measured values on the
measured value display.
Temperature mode
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).
Calibration
procedure
(only in pH
measuring mode)
CAL TEC AUTO
CAL CON 2P
CAL CON 1P
Simplified 2-point calibration using any
two different buffer solutions. The
nominal values of the buffer solutions
are stored in the sensor. This makes
the manual entry of the nominal values
redundant.
2-point calibration using the following
buffer solutions:
1.) pH 7.0 ± 0.5
2.) any pH value
The nominal values of the buffer
solutions must be entered.
1-point calibration using any single
buffer solution. The nominal value of
the buffer solution must be entered.
Calibration
(software version
2.18 or higher)
valid
invalid
last valid
active
abort
Displays and determines which calibration
data the measured value calculation is
based on. The active calibration is dis-
played in the calibration history (see sec-
tion 4.2.6).
valid indicates that a valid calibration is
available. The value cannot be changed.
invalid is displayed if the last calibration is
invalid and the sensor is blocked for mea-
surement. In this case, you can switch to
the values of the last valid calibration, pro-
vided a valid calibration is available in the
sensor. Thus you activate the last valid
calibration stored in the sensor the next
time you exit the setting table with Save
and quit. The next time the setting table is
opened, valid is displayed.