YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual
Page 17

SensoLyt
®
700 IQ (SW)
Commissioning
3 - 7
ba76006e01
01/2012
Carrying out settings
Using
s
, switch from the measured value display to the main menu of
the settings. Then navigate to the setting menu (setting table) of the
sensor. The exact procedure is given in the relevant IQ S
ENSOR
N
ET
system operating manual.
active indicates that the sensor is being
calibrated. If Cancel is selected, the active
calibration procedure is canceled as soon
as the setting table is exited with Save and
quit.
ORP shift
(only in mV
measuring mode)
-100 mV ... +100 mV
You can set the ORP zero point here.
Temperature
adjustment
-1.5 K ... +1.5 K
The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 K).
Notes:
Due to the thermal capacity of the
sensor, it is necessary to place it in a
container with at least 2 liters of water.
Leave the sensor in this container for at
least 15 minutes, or in the case of
temperature differences between the
water and sensor > 10 K for at least 1
hour, while stirring occasionally. Then
carry out the balancing procedure.
Save and quit
The system confirms the saving of the
settings and the display switches to the
next higher level.
Quit
The display switches to the next higher
level without saving the new settings.