6 technical data, 1 iq sensor netcomponentmiq/(a-)pr, Technical data – YSI IQ SensorNet MIQ/A-PR Module User Manual
Page 31: Chapter 6 t, Echnical, Data, S (see chapter 6 t, 6technical data
MIQ/(A-)PR
Technical data
31
ba76020e01
01/2012
6
Technical data
6.1
IQ S
ENSOR
N
ET
ComponentMIQ/(A-)PR
General technical data for the IQ S
ENSOR
N
ET
are given in the system
operating manual for the IQ S
ENSOR
N
ET
system.
Electrical data
Instrument safety
Terminal connections
Nominal voltage
Max. 24 VDC via the IQ S
ENSOR
N
ET
(for
details, see chapter T
ECHNICAL
DATA
of the
IQ S
ENSOR
N
ET
system operating manual).
Power consumption
3 W
Protective class
III
Number of MIQ/(A-)PR
modules in an
IQ S
ENSOR
N
ET
system
1
Applicable norms
– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1
IQ S
ENSOR
N
ET
connections
2
2 additional SENSORNET terminators that
can be activated (terminating resistors)
Terminal type
Screw-type terminal strip, accessible by
opening the lid
Terminal ranges
Solid wires:
Flexible wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
0.2 ... 2.5 mm
2
Cable feeds
4 cable glands M16 x 1.5 on the underside
of the module