3.2.4
PROFIBUS and module monitoring
The PROFIBUS module has 4 LEDs for monitoring the bus and one
LED for monitoring the PROFIBUS module. These 5 LEDs are
important aids for the monitoring and diagnosis of faults.
Bus monitoring
The LEDs have the following meaning:
Fig. 3-11 LEDs for bus monitoring
LED Color
Meaning
1
-
Not used
2
Green
The module is online (recognized by the
PROFIBUS master), data exchange is possible.
Off - The module is not online
3
Red
The module is offline, data exchange is not
possible.
Off - The module is not offline
4
Flashes
red
at a
freq. of:
1 Hz - error in the configuration:
During the initialization of the module, IN and/or
OUT data word length does not correspond to the
data word length in the configuration of the
PROFIBUS master (GSD file).
2 Hz - error in the user settings:
During the initialization of the module, the data
word length/content setting in the user settings
does not correspond to the data word length/
content setting in the configuration of the
PROFIBUS master (GSD file).
4 Hz - error during the initialization of the ASIC for
PROFIBUS communication.
Off - No error diagnostics available.