YSI IQ SensorNet MIQ/A-PR Module User Manual
Page 14
Interface for the PROFIBUS
MIQ/(A-)PR
14
ba76020e01
01/2012
(B)
Connecting the
MIQ/(A-)PR component
at the
PROFIBUS with a loop-
through
To loop the PROFIBUS cable through the MIQ/(A-)PR component, a
shielding terminal is included in the scope of delivery.
Fig. 3-7
Shielding terminal with PROFIBUS cable and shield connection cable
1
Screw
2
Washer
3
Insulating part of the shielding terminal
4
Cable insulation of the PROFIBUS cable
5
Shielding braid of the PROFIBUS cable
6
Wires of the PROFIBUS cable
7
Shield connection wire
1
Carefully strip the cable insulation of both PROFIBUS cables
for approx. 35 mm.
2
Shorten the exposed shielding braid to approx. 15 mm.
3
Bare the PROFIBUS wires for approx. 5 mm.
4
Feed both cables through the free cable glands into the
module housing of the MIQ/(A-)PR.
5
Create a shield connection wire (7). To do so, strip a wire on
both ends for approx. 5 mm and 10 mm.
6
Insert a washer in the insulating part (3) of the shielding
terminal.
7
Feed both PROFIBUS cables into the insulating part (3) and
lead them through the insulating part.
8
Feed the long stripped end of the shield connection wire (7)
into the insulating part (3) of the shielding terminal.
1
2
5
7
4
3
6