9 lists, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet MIQ/IC2 Module User Manual
Page 27: Lists -1, Explanation of the messages -1 9.1.1, Error messages -1, 9lists

MIQ/IC2
Lists
9 - 1
ba76034e01
01/2012
9
Lists
9.1
Explanation of the messages
In this chapter you will find a list with all the message codes and corre-
sponding message texts that may occur in the log book of the
IQ S
ENSOR
N
ET
system for the MIQ/IC2 current input module.
Note
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
All message codes of the MIQ/IC2 current input module end with the
number "381" (current input 1) or "382" (current input 2).
9.1.1
Error messages
Message code
Message text
EA9381
EA9382
Input current in undefined range
* Check settings and, if necessary, change them
* Check installation
* Check connected measuring system
* Check the MIQ/IC2
EAA381
EAA382
An error was reported
* Check settings and, if necessary, change them
* Check installation
* Check connected measuring system
* Check the MIQ/IC2
EI1381
Operational voltage too low
* Check installation and cable lengths,
Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
EI2381
Operational voltage too low, no operation possible
* Check installation and cable lengths,
Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components